UFL3 specifications

Tuneable titanium-sapphire oscillator with dispersion control

Coherent Chameleon Coherent Chameleon
Coherent OPO Compact Coherent OPO Compact
TUNABLE OSCILLATOR COHERENT CHAMELEON ULTRA II
Wavelength Range: 680 nm to 1080 nm
Tuning Speed: >40 nm/s
Power at peak (800nm): >2.8 mW
Power at 680 nm: >520 mW
Power at 700 nm: >1280 mW
Power at 920 nm: >1280 mW
Power at 1020 nm: >440 mW
Power at 1080 nm: >160 mW
Repetition Rate: 80 MHz, nominal
Astigmatism: <10%
Pointing (urad/nm): <0.5
OPO COHERENT OPO COMPACT
1000-1600nm signal IR automatic tuning >440mW at 1100nm
1750-4000nm idler IR automatic tuning >200mW at 1750nm
SHG Chameleon Auto-tracking SHG Units
Chameleon Auto-tracking SGH Input: 680-1080 nm
Output: 340-540 nm
Power at peak: ~1.6 W
Efficiency at peak: >45-50%
Chameleon OPO Compact SGH Input: 1060-1380 nm
Output: 530-690 nm
Power at peak: >130mW
Efficiency at peak: >30%
PULSE SELECTION PULSESELECT PULSE PICKER
Low Dispersion Design ***
Reflective Optic Configuration ***
Variable repetition rates Division ratio 1:20 - 1:5000
SiO2-Bragg Cell >50% Diffraction Efficiency
Low power (<2 W) optics sets 340 - 540nm
500 - 1600nm
High power version (> 2 W) 680-1080nm
SERVICE REQUIREMENTS ***
Electrical All units from 13 A single phase sockets
Optical Table A full sized optical table is recommended but the system can be supplied witha 112 mm thick metric breadboard (~100kg) and rigid support frame if required.
Diagnostics An autocorrelator and energy meter will be supplied.
Note on Laboratory Environmental conditions.

Prior to making an application for the system it is strongly recommended you contact LSF Staff.
For optimum operation of Amplifier, OPA, and mixing accesories, a draft free and +/- 1°C temperature controlled lab is the factory recommendation. It is anticipated that the complete system will perform over a range of +/- 3°C with perhaps some small drift in optimum performance. If your laboratory conditions fall outside of this range then please contact the LSF.

Notes

Laser goggles are not supplied with the system.

Host institution safety officers will be required to carry out a risk/hazard assessment of the equipment and its associated hazards before the equipment is installed.

For further information contact:

Dr Ian Clark
Central Laser Facility
STFC Rutherford Appleton Laboratory
Harwell Science & Innovation Campus
Didcot
Oxfordshire
OX11 0QX
Tel: +44 (0)1235 446 807

Calls for access are advertised on our calls for access page.